1. Performance Features
It is suitable for testing departments of large and medium-sized municipal hospitals;
5 point calibration, Detection range: Urinary iodine 20ug/L-1500ug/L;
Equipment calibration r value > 0.999;
No pre-digestion processing, automatic sample detection, automatic statistical analysis, data automatic upload;
Independent calibration, AI intelligent correction, bar code scanning, automatic generation of curves, quality control chart;
On-demand test function.
2.Technical Parameters
• Detection range: urine iodine: 20-1500 μg/L;
• Correlation coefficient: equipment calibration correlation coefficient (r) should be between 0.999 and 1.000;
• Linear range: within the range of (20 ~ 1500)μg/L, the linear correlation coefficient r≥0.990, the absolute deviation within the range of (20 ~ 100)μg/L is not greater than 15μg/L; The relative deviation is not greater than ±10% in the range of (100 ~ 1500)μg/L;
• Sensitivity: the minimum response value ≤20ug/L iodine content;
• Repeatability: Repeat the test with quality control (148 g/L) for 10 times, and the coefficient of variation (CV) is not greater than 6%;
• Batch difference: for reagents in different batches, test 10 times with quality control (148 g/L), coefficient of variation (CV) ≤5%;
• Fully automatic: users only need to add samples to be tested, and the whole analysis process is automatically completed by the instrument;
• Accuracy: When measuring the national urine iodine standard substance, the error of the measurement value is within its uncertainty range;
• Convenience: support for emergency insertion;
• Electromagnetic compatibility: shall comply with the requirements of GB/T18268.1-2010 and GB/T18268.26-2010.
3.General Parameters